Continuous and bimonthly publication
ISSN (on-line): 1806-3756

Licença Creative Commons
11752
Views
Back to summary
Open Access Peer-Reviewed
Artigo Original

Evaluation of manual resuscitators used in ICUs in Brazil

Avaliação de reanimadores manuais utilizados em UTIs brasileiras

Tatiana de Arruda Ortiz, Germano Forti Junior, Márcia Souza Volpe, Marcelo do Amaral Beraldo, Marcelo Britto Passos Amato, Carlos Roberto Ribeiro Carvalho, Mauro Roberto Tucci

ABSTRACT

Objective: To analyze psychological morbidity as a moderator of the relationship between smoking representations and quality of life in smokers and former smokers, as well as to determine which psychological variables discriminate between smokers with and without the intention to quit smoking. Methods: This was a quantitative, correlational cross-sectional study involving a convenience sample of 224 smokers and 169 former smokers. Results: In smokers and former smokers, psychological morbidity had a moderating effect on the relationship between mental/physical quality of life and smoking representations (cognitive representations, emotional representations, and comprehensibility). Smokers with the intention to quit smoking more often presented with low comprehensibility, threatening emotional representations, behavioral beliefs, and perceived behavioral control, as well as with normative/control beliefs, than did those without the intention to quit. Conclusions: The results of this study underscore the importance of the moderating effect exerted by psychological morbidity, as well as that of sociocognitive variables, among smokers who have the intention to quit smoking.

RESUMO

Objetivo: Analisar a morbidade psicológica como um moderador na relação entre as representações do tabaco e a qualidade de vida em fumantes e ex-fumantes, assim como conhecer as variáveis psicológicas que discriminam os fumantes com e sem intenção para deixar de fumar. Métodos: Estudo quantitativo, correlacional e transversal com uma amostra de conveniência constituída por 224 fumantes e 169 ex-fumantes. Resultados: Verificou-se um efeito moderador da morbidade psicológica na relação entre a qualidade de vida (física e mental) e as representações do tabaco (representações cognitivas e emocionais e compreensão) nos fumantes e nos ex-fumantes. Os fumantes com intenção para deixar de fumar apresentavam menor compreensão, representações emocionais mais ameaçadoras, mais crenças de comportamento, maior controle comportamental percebido e mais crenças normativas/controle do que aqueles sem essa intenção. Conclusões: Os resultados deste estudo enfatizam a importância da morbidade psicológica como moderadora, bem como das variáveis sociocognitivas, junto dos fumantes que querem deixar de fumar.

Palavras-chave: Hábito de fumar/psicologia; Abandono do hábito de fumar; Hábito de fumar/prevenção & controle.

Introduction

Manual resuscitators (MRs) are used for manual ventilation of patients. They are also known as "ambu", which is the trademark of a resuscitator and the term used as a synonym for MR by professionals in Brazil and other countries.(1,2) Being used in ICUs in order optimize oxygenation and facilitate the removal of secretion in patients on invasive mechanical ventilation, MRs are also used for intra-hospital transport and as a source of temporary ventilation in intubated and nonintubated patients.(1-4) On the basis of their inflation mechanism, MRs can be classified as self-inflating or flow-inflating,(1,5) the former being the most commonly used type in ICUs in Brazil.


Self-inflating MRs, which is the type of MR evaluated in the present study, consist of a compressible self-inflating bag, an inflating valve, and an inspiratory/expiratory valve (patient valve; Figure 1).(5,6) When the bag is compressed by the operator, the patient valve occludes the expiratory port and allows the air within the bag to flow to the patient; when the pressure within the bag ceases, the patient valve returns to its normal position, occluding the communication of the bag with the patient and allowing the air coming from the patient to be exhaled through the expiratory port. A drop in pressure within the bag during exhalation causes the inflating valve to open, allowing the air to enter the bag.(5,6) Additional components of this type of MR include(5,6) a device for enriching the oxygen mixture, a positive end-expiratory pressure (PEEP) valve, and a pressure-limiting valve (PLV), the use of which in adult patients is not in accordance with the American Society for Testing and Materials (ASTM) criteria.(7)



In the literature, there are studies evaluating the performance and safety of MRs available in other countries in accordance with international technical standards.(8-10) Comparative studies allow the identification of problems and the improvement of these devices. The objective of the present study was to evaluate the characteristics and performance of some of the self-inflating MRs for use in adult patients that are available in the Brazilian market, in accordance with international standards.

Methods

The present study was approved by the Research Ethics Committee of the University of São Paulo School of Medicine Hospital das Clínicas and was carried out in the Experimental Mechanical Ventilation Laboratory of the Department of Pulmonology of the same institution in 2006.

We evaluated eight MRs, of which five were manufactured in Brazil (J.G. Moriya, São Paulo; Oxigel, São Paulo; Protec Equipamentos Médico-Hospitalares, São Paulo; RWR Equipamentos Hospitalares, São Bernardo do Campo; and Unitec Ind. e Com. de Aparelhos Hospitalares Ltda., São Paulo), two were manufactured in Denmark (Ambu® Silicone Resuscitator and Ambu® SPUR; Ambu A/S, Ballerup, Denmark), and one was manufactured in the USA (Hudson RCI® Lifesaver; Hudson RCI, Durham, NC, USA). The MRs tested were lent by the manufacturers or their representatives in Brazil. All of the models are still being marketed, and the updates of the MRs since the present study was carried out to this writing are described in Table 1.



Since there are no Brazilian standards defining the characteristics for the manufacture and marketing of MRs, the present evaluation used the ASTM criteria, Designation F920-93,(7) which are equivalent to the International Organization for Standardization 8382 criteria,(11) used in other countries. Using a respiratory system simulator, the following criteria were evaluated: resistance imposed by the patient valve during the inspiratory and expiratory phases; functioning of the PLV; and tidal volume (VT) delivered by the MR. In accordance with the ASTM criteria, the following specifications should be met: a drop < 5 cmH2O in inspiratory and expiratory pressure in the presence of a flow of 50 L/min, leading to resistance < 6 cmH2O . L−1 . s−1; the use of a PLV is not mandatory for MRs for adults, but, if this valve is available, it should be possible to close it; a VT  600 mL should be delivered to adult patients (weight  40 kg), with resistance of 20 cmH2O . L−1 . s−1 and compliance of 0.02 L/cmH2O, using the one-handed technique (maximum hand size, 185 mm).

The measurement system consisted of a plastic tube to which pressure and flow sensors were connected (Figure 2). The signals from the sensors were amplified and stored on a computer for subsequent analysis with the Labview® software (National Instruments, TX, USA). The signal sampling frequency was 200 Hz.



In order to measure inspiratory and expiratory resistance, a mechanical ventilator (Bear 5®; Bear Medical Systems, Riverside, CA, USA) was connected to one end of the plastic tube with the sensors (Figure 2). The ventilator was set to the controlled volume mode, with a constant inspiratory flow of 60 L/min and a PEEP of zero cmH2O. In the evaluation of expiratory resistance, the patient connection port of the MR was attached to the other end of the plastic tube, and the air from the ventilator left through the expiratory port into room air (Figure 2). In the evaluation of inspiratory resistance, the MR was disassembled and only the proximal portion of the set, without the bag, was used (Figure 2). The part of the MR that is connected to the bag was attached to the other end of the plastic tube, and the air left through the patient connection port to the environment. The PLV, when present, was occluded. The Ambu® Silicone Resuscitator was not tested because it could not be disassembled. The two resistances were calculated as the difference between the proximal and distal pressure to the valve (distal pressure was not measured, because it was equal to zero, since the system was open to the environment), divided by the measured flow.

For the testing of VT, the MRs were connected to the plastic tube with the sensors, and this tube was connected to a tracheal cannula of 8.5 mm in diameter (Rusch, Montevideo, Uruguay), the distal portion of which, with the cuff inflated, was fitted into another plastic tube that was connected to a respiratory system simulator (TTL 2600®; Michigan Instruments©, Grands Rapids, MI, USA; Figure 2). The simulator allowed the compliance of the bellows to be set with springs; resistance was set with nonlinear resistors.

Four volunteer ICU physiotherapists in the city of São Paulo (two males and two females) operated the MRs. The size of the dominant hand (from the wrist to the tip of the middle finger) was 200 mm for male1, 188 mm for male 2, 175 mm for female 1, and 160 mm for female 2. The physiotherapists were instructed to inflate the model as if they were ventilating a patient, first using the one-handed technique and then using the two-handed technique, with the eight MRs, which were tested in a random sequence, under two conditions of resistance and compliance: 1) "ASTM" condition-compliance of 0.02 L/cmH2O and resistance of 20 cmH2O . L−1 . s−1, in accordance with the ASTM criteria;(7) and 2) "normal" condition-compliance of 0.05 L/cmH2O and resistance of 5 cmH2O . L−1 . s−1, simulating a patient without lung disease.

In order to calculate VT, three respiratory cycles were analyzed, all of which were randomly selected among the recorded cycles (a minimum of 10 per test condition). Since each volunteer (totaling four) handled eight MRs, using the one-handed and two-handed techniques, in two distinct clinical settings (ASTM condition and normal condition), 128 tests were performed.

Of the eight MRs, five were equipped with a PLV and were submitted to this test. The ventilator, connected to one end of the measurement system (with a flow sensor and a pressure sensor), was set to the controlled volume mode, with an inspiratory flow of 20 L/min, a VT of 1,800 mL, and a PEEP of zero cmH2O. A part of the MR was connected to the other end, as were another (distal) flow sensor and the simulator, in sequence (Figure 2). The pressure alarm of the mechanical ventilator was deactivated. We determined the pressure at which the PLV opened and allowed air leaks (a time point that could be identified when the flow in the distal sensor decreased in relation to that in the proximal sensor), this pressure being referred to as "opening pressure", and determined the pressure at which all air sent by the ventilator passed only through the PLV (a time point at which the distal flow was equal to zero), this pressure being referred to as "maximal pressure". The functioning of the valve was considered appropriate if the opening pressure was within the pressure interval suggested by the manufacturers, in general, values greater than 35-40 cmH2O (Table 1).

For each MR, only one inspiratory and expiratory resistance value was shown, because the variation between the cycles was negligible. The testing of the PLV was performed only once per MR. The VT values were expressed as mean and standard deviation and were evaluated by multivariate analysis using a mixed linear regression model for each of the four conditions studied: VT generated with the one-handed technique under the ASTM and normal conditions; and VT generated with the two-handed technique under the same two conditions. The physiotherapists and the MRs were evaluated as fixed effects. Post hoc tests with Sidak correction were used for comparisons of the VT values among the physiotherapists and among the MRs. For this analysis, we used the Statistical Package for the Social Sciences, version 17 for Windows (SPSS Inc., Chicago, IL, USA). Values of p < 0.05 were considered significant.

Results

Table 1 shows the characteristics of the MRs tested. The most common type of patient valve was a diaphragm, and the most common bag material was silicone. The capacity of the bag of each MR varied (1,000-2,500 mL). The three imported MRs did not have a PLV, which was a feature of the five Brazilian-made MRs, and, in the MR manufactured by RWR, the PLV could be closed without there being any leaks. Another difference observed between the imported and the Brazilian-made MRs was that the patient connection port was fixed in the five Brazilian-made MRs, whereas, in the imported ones, this connection port was a "360° swivel".

Table 2 describes the results of the tests evaluating the resistances of the patient valve during inhalation and exhalation, as well as the functioning of the PLV. Only in the Protec RM did expiratory resistance exceeded the limit set by the ASTM criteria (< 6 cmH2O . L−1 . s−1). In the other MRs tested, the resistances met those criteria, the highest inspiratory resistance being found in the Oxigel MR (5.69 cmH2O . L−1 . s−1). The PLV, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O.



The mean delivered VT values, by MR (Table 3) and by physiotherapist (Table E1, available online at http://www.jornaldepneumologia.com.br/detalhe_artigo.asp?id=2212), varied. In the situations evaluated (handling of the eight MRs, with the use of the one-handed and two-handed techniques, in two distinct clinical settings-ASTM condition and normal condition), there were significant differences (p < 0.001) among the MRs and among the physiotherapists. All of the MRs, whenever the one-handed technique was used under the ASTM condition, delivered a lower VT than that recommended by the ASTM ( 600 mL). Only the two models of MR manufactured by Ambu A/S delivered a VT  600 mL when the one-handed technique was used, but only during the simulation of the normal clinical condition. Even when the two-handed technique was used, the MRs manufactured by Oxigel and RWR did not deliver the minimal VT under the ASTM clinical condition, whereas the MRs manufactured by Protec and Unitec, regardless of the clinical setting simulated, delivered a VT < 600 mL.



In addition, when we observe the results reported in Table 3, we can see that the VT values generated under the normal clinical condition were, in general, higher than were those generated under the ASTM condition, which simulates a pulmonary condition with lower compliance and higher resistance, and that most of the VT values obtained with the MRs that did not have a PLV were higher than were those obtained with the MRs that had a PLV.

Discussion

The main findings of the present study can be summarized as follows: 1) the (inspiratory and expiratory) resistances of the patient valve met the ASTM criteria in all but one of the MRs tested; 2) the eight RMs evaluated did not deliver the minimal VT under conditions standardized by those same criteria (VT  600 mL, with the use of the one-handed technique, under the ASTM clinical condition); 3) the use of the two-handed technique did not solve the problem in four MRs (VT remained < 600 mL); and 4) the PLV present in the Brazilian-made MRs usually led to the delivery of a lower VT than that delivered by the MRs that did not have a PLV.

In Brazil, there have been few published studies evaluating MRs for adult patients, commonly used in ICUs in the country.(5,12,13) Although those studies investigated important characteristics, such as the oxygen fraction delivered by MRs with and without an oxygen reservoir, the present study evaluated other equally important properties that had not yet been reported in the national literature.

Regarding the evaluation of resistance of the patient valve during inhalation and exhalation, all of the MRs tested were within the specifications of the ASTM, except for the Protec MR, in which expiratory resistance was 7.91 cmH2O . L−1 . s−1. This result is not surprising, given that international studies had described MRs in which the resistances of the patient valve were higher than recommended.(14,15) It is of note that higher expiratory pressure translates to longer exhalation, which can cause air trapping (auto-PEEP) and risk of barotrauma,(15) especially when the MR is handled at a high respiratory rate and a high VT.

For the evaluation of the VT generated with the one-handed and two-handed techniques, we selected male and female volunteers, because it is known that the VT obtained varies with the gender of the operator (VT is lower for female operators).(16) Mean VT varied greatly among the MRs tested, a finding that had been reported in other studies.(17) We observed that some Brazilian-made MRs delivered a low VT even with the use of the two-handed technique, which generates a higher VT than does the one-handed technique.(2,17,18) Other authors had reported, even in normal compliance simulations, VT values lower than 600 mL.(17) However, in the study by Mazzolini et al.,(10) which evaluated 16 disposable MRs manufactured abroad, only three MRs did not deliver the minimal VT recommended by the ASTM ( 600mL).

It is possible that various factors influenced the results for VT. One factor that seems to have led to increased VT values was the size of the hands of one of the physiotherapists (200 mm, a larger size than that recommended by the criteria, Male 1, Table E1, available online at http://www.jornaldepneumologia.com.br/detalhe_artigo.asp?id=2212). This association between delivered VT and hand size has been described by other authors.(6,8,17,18) Factors that could lead to low volumes are the size(18) and the material of the bag of the resuscitator tested.(8) The volume of air in the bag is not totally delivered to the patient when the bag is compressed, and, in general, the volume of the bag should be twice the volume to be delivered.(8) A bag made of a material that is not very compliant (such as polyvinyl) can lead to a lower VT.(8) In our study, the MR that delivered the lowest VT (Unitec) had the bag with the lowest volume, and, in addition, this bag was made of polyvinyl. It is of note that VT was generally lower from the Brazilian-made MRs than from the imported ones, except for the MR manufactured by RWR, which generated a higher VT, probably because the capacity of the bag was 2,500 mL.



One could question whether the low VT values found for some of the MRs tested resulted from fatigue of the volunteers. However, the study by Hess et al.,(16) which evaluated the effect of fatigue caused by ventilation with a MR for 30 min under low compliance (20 mL/cmH2O), found no reduction in VT after 30 minutes for the 14 volunteers evaluated. Considering that, in our study, total test time for each of the MRs was approximately one minute, which would total eight minutes for each of the four situations evaluated (two impedance values with the use of the one-handed and two-handed techniques), and that there was a rest break between each measurement, it seems unlikely that the occurrence of fatigue would have influenced the results.

As expected, the VT obtained in the simulation of a "normal" patient was higher than that found in the tests using the compliance and resistance settings established by the ASTM criteria, which simulated a patient with decreased compliance and increased resistance, which are proven to lead to a lower VT.(17)

The presence of a PLV in the Brazilian-made MRs possibly impaired their performance. All of the Brazilian-made MRs had a PLV, and in only one of them could the valve be closed. The valves of all of them initially opened at low pressures, the values of which were below the specified value, causing leaks during inflation, which can partially explain the lower VT obtained with the RMs that had a PLV. This result can be considered a severe fault, because it can prevent appropriate ventilation in patients with decreased lung compliance or with increased airway resistance, such as those with pulmonary edema and those with asthma attacks, respectively.

According to some authors, the presence of PLV does not bring additional safety to adult patients,(9) given that the risk of barotrauma is not associated only with high peak pressures during inhalation.(19) In fact, the ASTM criteria(7) dispense with the use of a pressure-regulating device for MRs for adults and determine that, if this valve is present, it should be possible to lock it and its closing mechanism should be quite evident to the operator.(6,7) Even in MRs for neonatal/pediatric patients, for which the presence of this valve is recommended in order to limit pressure to values below 40 ± 5 cmH2O,(7,20) malfunctioning of the PLV can also impair ventilation.(2)

We consider limitations of our study the fact that all tests were performed using only one unit of each MR (however, all MRs were brand-new); that resistance of the patient valve was tested at a flow of 60 L/min, rather than of 50 L/min (value recommended by the ASTM), which can hinder the comparison with other studies; and that, since the present study was carried out in 2006, some MR models underwent changes (Table 1), which might have altered their performances. This should be probably true for the MR manufactured by Unitec, in which the capacity of the bag was significantly increased (Table 1).

Despite the limitations of this study, we believe that important deficiencies of the MRs manufactured in Brazil are reported, especially those regarding the functioning of the PLV (most could not be locked and opened early), deficiencies that, in some clinical conditions, can impair ventilation in critically ill patients. Our results indicate that, in order to deliver an appropriate VT, it is necessary that both hands be used to compress the MR bag and that, when allowed by the MR, the PLV be kept closed. We believe that the results obtained in the present study can aid in performing future tests and in improving the Brazilian-made MRs.

References

1. White GC. Emergency resuscitation equipment. In: White GC, editor. Equipment Theory for Respiratory Care. Clifton Park: Thomson/Delmar Learning; 2005. p. 249-99.

2. Oliveira PM, Almeida-Junior AA, Almeida CC, Ribeiro MA, Ribeiro JD. Fatores que afetam a ventilação com o reanimador manual autoinflável: uma revisão sistemática. Rev Paul Pediatr. 2011;29(4):645-55.

3. Maxwell L, Ellis ER. The effects of three manual hyperinflation techniques on pattern of ventilation in a test lung model. Anaesth Intensive Care. 2002;30(3):283-8. PMid:12075634

4. Stiller K. Physiotherapy in intensive care: towards an evidence-based practice. Chest. 2000;118(6):1801-13. http://dx.doi.org/10.1378/chest.118.6.1801 PMid:11115476

5. Godoy AC, Vieira RJ, Vieira Neto RJ. Oxygen outflow delivered by manually operated self-inflating resuscitation bags in patients breathing spontaneously. J Bras Pneumol. 2008;34(4):212-6. PMid:18425257

6. Dorsch JA, Dorsch SE. Manual Resuscitators. In: Dorsch JA, Dorsch SE, editors. Understanding Anesthesia Equipment. Baltimore: Williams & Wilkins; 1994. p. 225-38.

7. ASTM International. ASTM Standard F920-93 Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans. West Conshohocken: ASTM International; 1999.

8. Hess DR. Manual and Gas-Powered Resuscitators. In: Branson RD, Hess DR, Chatburn RL, editors. Respiratory Care Equipment. Philadelphia: Lippincott; 1999. p. 187-204. PMid:10369141

9. Eaton JM. Medical technology. Adult manual resuscitators. Br J Hosp Med. 1984;31(1):67-70. PMid:6697045

10. Mazzolini DG Jr, Marshall NA. Evaluation of 16 adult disposable manual resuscitators. Respir Care. 2004;49(12):1509-14. PMid:15571642

11. International Organization for Standardization. International Standard ISO 8382: 1988(E). Resuscitators intended for use with humans. New York: American National Standards Institute; 1988.

12. de Godoy AC, Vieira RJ. Comparison of the FiO2 delivered by seven models of the self-inflating bag-mask system. Rev Bras Anestesiol. 2009;59(1):21-7. PMid:19374212

13. Godoy AC, Vieira RJ, De Capitani EM. Alterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate. J Bras Pneumol. 2008;34(10):817-21. http://dx.doi.org/10.1590/S1806-37132008001000010 PMid:19009215

14. Barnes TA, McGarry WP 3rd. Evaluation of ten disposable manual resuscitators. Respir Care. 1990;35(10):960-8. PMid:10145334

15. Hess D, Simmons M. An evaluation of the resistance to flow through the patient valves of twelve adult manual resuscitators. Respir Care. 1992;37(5):432-8. PMid:10145632

16. Hess D, Simmons M, Blaukovitch S, Lightner D, Doyle T. An evaluation of the effects of fatigue, impedance, and use of two hands on volumes delivered during bag-valve ventilation. Respir Care. 1993;38:271-5.

17. Hess D, Spahr C. An evaluation of volumes delivered by selected adult disposable resuscitators: the effects of hand size, number of hands used, and use of disposable medical gloves. Respir Care. 1990;35:800-5. PMid:10145317

18. Hess D, Goff G, Johnson K. The effect of hand size, resuscitator brand, and use of two hands on volumes delivered during adult bag-valve ventilation. Respir Care. 1989;34:805-10.

19. Ricard JD. Manual ventilation and risk of barotrauma: primum non nocere. Respir Care. 2005;50(3):338-9. PMid:15779151

20. Zaconeta CAM, Borges MBS, Souza DVB, Marques MG. Avaliação do pico de pressão e da frequência respiratória durante o uso de balão autoinflável por socorristas do Corpo de

21. Bombeiros em um modelo de pulmão neonatal pré-termo. Rev Paul Pediatr. 2010;28(1):5-9. http://dx.doi.org/10.1590/S0103-05822010000100002




* Study carried out in Laboratório de Investigação Médica 09 (LIM09, Laboratory for Medical Research 09), specializing in Pulmonology, University of São Paulo School of Medicine, São Paulo, Brazil.
Correspondence to: Mauro R. Tucci. Laboratório de Pneumologia LIM09, Faculdade de Medicina da Universidade de São Paulo, Avenida Dr. Arnaldo, 455, Sala 2144, 2º andar, CEP 01246-903, São Paulo, SP, Brasil.
Tel. 55 11 3061-7361. Fax: 55 11 3061-2492. E-mail: mrotucci@gmail.com
Financial support: This study received financial support from the Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP, São Paulo Research Foundation), the Laboratory for Medical Research of the University of São Paulo School of Medicine Hospitals das Clínicas, and the Financiadora de Estudos e Projetos (FINEP, Brazilian Financing Agency for Studies and Projects).
Submitted: 21 February 2013. Accepted, after review: 6 September 203.




About the authors

Tatiana de Arruda Ortiz
Physiotherapist. Laboratório de Investigação Médica 09 (LIM09, Laboratory for Medical Research 09), specializing in Pulmonology, University of São Paulo School of Medicine, São Paulo, Brazil.

Germano Forti Junior
Physiotherapist. Laboratório de Investigação Médica 09 (LIM09, Laboratory for Medical Research 09), specializing in Pulmonology, University of São Paulo School of Medicine, São Paulo, Brazil.

Márcia Souza Volpe
Adjunct Professor. Department of Applied Physiotherapy, Universidade Federal do Triângulo Mineiro, Uberaba, Brazil.

Marcelo do Amaral Beraldo
Physiotherapist. Laboratório de Investigação Médica 09 (LIM09, Laboratory for Medical Research 09), specializing in Pulmonology, University of São Paulo School of Medicine, São Paulo, Brazil

Marcelo Britto Passos Amato
Supervising Physician. Respiratory ICU, University of São Paulo School of Medicine Hospital das Clínicas, São Paulo, Brazil.

Carlos Roberto Ribeiro Carvalho
Full Professor of Pulmonology. University of São Paulo School of Medicine, São Paulo, Brazil.

Mauro Roberto Tucci
Physician. Laboratório de Investigação Médica 09 (LIM09, Laboratory for Medical Research 09), specializing in Pulmonology, University of São Paulo School of Medicine; and A.C. Camargo Cancer Hospital, São Paulo, Brazil.

Indexes

Development by:

© All rights reserved 2024 - Jornal Brasileiro de Pneumologia