Nancy Spekla Grande, Romilda Aparecida Nakayama,
Antonia Maria de Oliveira Machado, Fábio Akio Yamaguti, Cesar Uehara
J Bras Pneumol.2002;28(5):250-260
Introduction: Bronchoscopic is a diagnostic and therapeutic procedure performed by introducing a flexible tube in the airways that allows for the visualization even of the smallest airways. In order to prevent or reduce the risk of infection, the bronchoscope must be adequately disinfected with at least high level disinfection. Purpose: Check for the risk of bacterial contamination for patients submitted to bronchoscopy at the Hospital São Paulo of Unifesp University, State of São Paulo. Methods: From 1997 to 1998, bronchoscope reprocessing included cleaning, rinsing with sterile or potable water, followed by rinsing with 2% glutaraldehyde for 20 minutes, rinsing again with sterile or potable water, and rinsing with 70% ethyl alcohol, and forced-air drying through the suction channel. Samples of 65 patients were collected for microbiologic tests by instilling sterile saline solution through the suction channel of the bronchoscope. Results: After reprocessing, the following were found in the samples: Staphylococcus epidermidis, Enterobacter sp, Acinetobacter baumanni, Streptococcus viridans, Staphylococcus aureus, Streptococcus beta hemoliticus A, Staphylococcus coagulase negative, and mycobacterial growth was found in the culture of five samples. Conclusions: Bronchoscope disinfection with 2% glutaraldehyde was not sufficient to assured disinfection of the scope and the presence of Staphylococcus epidermidis shows that there was bronchoscope contamination due to handling after disinfection.